Product Name: |
Automatic COD Analyser. |
Model: |
COD-201. |
Measurement Object: |
COD concentration of water. |
Measurement Method: |
100°C digestion with KMnO4 acidified by sulfuric acid (in conformity with JIS K0102) or 100°C digestion with alkaline KMnO4.
|
End Point Detection: |
Oxidation / reduction potentiometric titration.
|
Measurement Range: |
Single or dual range (units: mg/l) Specify one of the following:- |
(1) Single Range: |
0~20, 0~30, 0~40, 0~50, 0~100, 0~200, 0~300, 0~400, 0~500, 0~1000.
|
(2) Dual Range: |
0~20/0~40, 0~30/0~60, 0~40/0~80, 0~50/0~100, 0~100/0~200, 0~200/0~400, 0~300/0~600, 0~400/0~800, 0~500/0~1000 (automatic range selection).
|
(3) 2 ranges at 2 points: |
2 ranges can be selected from those listed in (1) above (range at each measurement point is fixed).
|
Measurement Cycle: |
1 hour per measurement, intermittent measurement cycles available (1,2,3,4,5,6, hour settings). |
Measurement Points: |
Single or dual points. |
Display: |
Digital, touch-screen. |
Recording Method: |
Printer. (1) Year, month, day, hour, minutes. (2) Measurement value mg/l. (3) Measurement status. (4) Analyser fault code. (5) Daily report (max., min., average value, mg/l). |
Input Signals: |
Remote Start Signal: |
Closed contact point every hour. |
Monitoring Station Stop: |
(Telemetry failure) open contact indicates failure. |
Output Signals: |
Measurement Value: |
Isolated 4~20mA (Max. 600W load). |
Under Maintenance: |
Closed contacts during maintenance. |
Analyser Fault Signals: |
Closed contacts if any of the following is detected. (1) Power failure (2) Low cleaning water (3) Low reagent (4) Low sample flow (5) Temperature fault (6) Sensor failure |
Temperature & Humidity: |
5~40°C, less than 85% (RH). |